novembre 2007
Digital systems aim to reduce breast imaging radiation dose
Photon-counting, high DQE, x-ray tubes, antiscatter grids, and 3D imaging all offer strategies that may lower radiation exposure
Because modern digital detectors can be tuned to be more efficient
than screen-film systems, they automatically lower the radiation dose
to the patient while maintaining the image quality radiologists crave.
Photon-counting, the method used in Sectra's detector, for example,
minimizes patient exposure to radiation. In a 2006 analysis, the
British National Health Service Breast Screening Programme reported
that the Sectra MicroDose Mammography system achieved significant dose
savings over screen-film mammography. The average mean glandular dose
of the MicroDose System in this analysis was 0.57 mGy, a level that was
75% lower than average dose levels in film imaging systems.
A Swedish study1 confirms that clinical performance in terms of cancer
detections rate and recall rate can be improved, while decreasing the
radiation dose.
"The secret to the dose reduction possible with Sectra MicroDose
Mammography is a high detective quantum efficiency (DQE) and contrast
transfer, which basically translates to very low noise in the system
and absorption and use of all incident x-rays with high efficiency,"
said Prof. Mats Danielsson, a professor of physics at the Royal
Institute of Technology in Stockholm and cofounder of Sectra Mamea.
"The photon-counting system, in fact, rejects all electronic noise that
would otherwise blur the image."
Manufacturers of digital mammography systems are seeking to reduce
radiation dose to patients by other means as well. One approach
involves changing the way in which x-rays are generated. By going to a
higher kV, the average x-ray energy will be higher, which means it will
penetrate the breast and hit the detector more efficiently, said Dr.
Andrew Smith, principal scientist for Hologic of Bedford, Massachusetts.
"For the same amount of signal hitting the detector, there is less
signal being absorbed in the breast," Smith said. "That is something
that could not be done with screen-film imaging because screen-film
systems must operate in a very limited kiloelectron volt range to
provide a high-quality image. But digital detectors are capable of
imaging over a wide range of exposures and incident kV energies. So
changing x-ray generation is the strategy by which people are further
reducing dose."
The focus of change is on composition of the x-ray tube and x-ray
filters. The gold standard for analog imaging has been a molybdenum
x-ray tube and a molybdenum x-ray filter. However, the x-rays emitted
from a tungsten tube have higher energy.
"Tungsten generates higher energy, which penetrates the breast better,
so less radiation is absorbed in the breast and more goes to the
imaging plate itself," said Jonny Eser, mammography product manager in
the U.S. for Siemens. "Because of the higher current used to excite the
electrons, there is better penetration-particularly in dense
breasts-and a reduction in the patient dose."
The most common molybdenum form of x-ray tube operates at approximately
35 kV constant potential and produces peaks of radiation at 18 keV and
20 keV. That is the narrow x-ray spectrum of a molybdenum anode target
and filter.
X-ray tubes with tungsten targets operate at 30 kV, and they generate a
large proportion of characteristic radiation at 8 keV and 10 keV. In
combination with a rhodium filter, both molybdenum and tungsten x-ray
tubes improve the depiction of glandular tissue, pectoral muscle, skin,
and subcutaneous tissue for all types of breasts. The tungsten/rhodium
combination provides a better depiction of glandular tissue in dense
breasts, however, according to various studies.
Siemens Medical Systems has used tungsten tube technology in its
mammography units since it first released the Mammomat 3000 Nova analog
product line. In phantom studies2,3 of its amorphous selenium
full-field digital Mammomat Novation system, investigators have shown
that tungsten/rhodium preserves image quality while reducing radiation
dose. Dose reductions ranged from 28% for an average 4-cm breast to 66%
for a thick, fatty breast. Signal-to-noise ratios in average breasts
nevertheless increased by 28% for calcifications and 48% for masses,
and they were up to three times higher in fatty breasts, in a study by
Duke University3 in Durham, North Carolina.
A 2006 European study4 found that the mean glandular dose for the
tungsten/rhodium combination used in the Novation mammography system
was 49% lower than the molybdenum/molybdenum tube/filter configuration
and 33% lower than the molybdenum/rhodium combination.
Sectra has a rotating tungsten anode, and Hologic is planning to add a
tungsten alternative to its Selenia system.
Although Siemens recommends using tungsten for imaging breasts of all
sizes and tissue compositions, its digital mammography system offers
both molybdenum and tungsten to suit customers' preferences.
GE Healthcare's digital mammography system has an x-ray tube with a
rhodium as well as a tungsten track. Depending on breast density and
compression thickness, the filtration of the x-ray system can be
changed by means of automatic optimized protocols, which adjusts the
energy spectrum of the x-ray beam to try to get the best dose
performance, said Mike Barber, the company's chief technology officer.
"The automatic optimized protocols determine-based on a number of
parameters such as kV or mA settings as well as breast compression
thickness-what is the right energy spectrum," Barber said. "And the way
you get that different energy spectrum is by switching between rhodium
and tungsten."
Dose might be reduced further, at least theoretically, by eliminating
the scatter-absorbing grid. This grid has long been a standard element
in mammography and radiography to counteract the loss of contrast due
to scattered radiation, according to Eser.
Grids have been especially important in mammography because of the need
for high contrast to distinguish between malignant and normal tissues,
which often have nearly the same densities.
After researchers removed the antiscatter grid in one study,5 image
quality improved when breast phantoms were smaller than 5 cm, and
contrast detail with and without the grid was similar when breast
thickness was between 5 cm and 7 cm, leading investigators from the
University Medical Center of Nijmegen in the Netherlands to suggest
that a grid might not be necessary when imaging compressed breast
thicknesses of 1 to 7 cm.
Researchers from Siemens reported last year6 that an antiscatter grid
improved image quality only when breast tissue was 5 cm or greater. The
investigators recommended removing the grid for imaging breast
thicknesses less than 5 cm because the amount of scatter radiation in
such tissue samples is low.
Danielsson contends, however, that the research does not indicate that
removing the grid leads to dose reduction for average or thicker
breasts.
"Instead, the situation, from image quality and dose point of view, is
equally bad in both cases, with or without the grid, and far away from
the performance of the Sectra multislit scanning technique," he said,
citing a peer-reviewed paper.7
A multislit system, such as the Sectra MicroDose, maintains the
contrast that would be reduced by use of a bucky grid and still rejects
scattered radiation without compromising primary radiation. As a
result, it operates at radiation doses below those associated with
detectors that have a grid. Removing the grid for those detectors would
not reduce dose for an average breast since the fraction of scattered
radiation would increase.
Research published in the April 2006 issue of Medical Physics
summarizes the effect of removing the grid. For thinner breasts, its
removal is advantageous, while for thicker breasts, the penalty in
terms of increased scatter is too high. In all cases, the performance
of the scanned-slit photon-counting technique was found to be superior
to the other techniques, according to the investigators.
GE is working to further improve dose performance by increasing the
sensitivity of the amorphous silicon flat-panel detector to leverage
every photon that reaches the scintillator, according to Barber, who
could not divulge the specific strategies the company is following at
this point.
Future dose-reduction strategies may not have much to do with digital
detector technology, according to Smith.
"With selenium detector technology, we are already stopping almost 100%
of the x-rays, so there's not much to do in terms of making the
detector more radiation-efficient," he said.
Absorption, however, is only one of several factors affecting DQE and,
thus, dose and image quality. In terms of noise, there is much room for
improvement.
In a study conducted by researchers from Lausanne, Switzerland, DQE for
all available technologies for digital mammography was measured, and it
was found that, depending on detector, improvements of 40% to 80% are
possible. At low doses, the potential improvement is even higher.8
Sectra's photon-counting technique comes out on top in this comparison,
and the reason is the absence of electronic noise.
The way to get improvement, according to Smith, is by moving to 3D
imaging. Volumetric data capture through tomosynthesis does not improve
the dose efficiency of the digital detector, nor does it improve the
resolution of images. But it does provide a clearer depiction of
anatomic structures and abnormalities by expanding beyond the standard
2D views of the breast.
Tomosynthesis also may contribute to dose reduction by requiring only
two views rather than the four views of each breast in a standard
screening exam, Eser said.
"There need to be some studies to actually prove that you could
eliminate two of the regular views with tomosynthesis," he said. "But
if it proves true, that will also be a reduction of dose to the
patient."
A potential breakthrough for the future is the introduction of active
x-ray filters in terms of x-ray optics fabricated using advances in
nanotechnology, according to Danielsson.
"These active lenses would shape the x-ray beam to enable further dose
reduction by tuning it to the optimum energy, an energy that could
potentially be changed for different breast sizes," he said. "Research
is actively going on in this area at a number of universities and
companies, but it is hard today to guess when this might become a
clinical reality."
It is interesting to note, however, that mammography, the last field to
switch from screen-film to digital, is now already using
photon-counting systems in clinical praxis, a technique that is still
in the laboratories for use in next-generation CT systems, Danielsson
said.
Ms. Sandrick is a freelance writer based in Chicago.OTHER METHODS